Validation result

Information

File Name
00000001.dcm
OID :
1.3.6.1.4.1.12559.11.11.4.33543
Validation Date :
12/1/25 9:23:31 AM (CET GMT+0100)
PixelMed (Wed Dec 18 15:38:40 EST 2019)
DICOM SR Conformance (Wed Dec 18 15:38:40 EST 2019)
Validation Results :
DONE_FAILED
Permanent link :
https://connectathon.ihe-catalyst.net/evs/report.seam?oid=1.3.6.1.4.1.12559.11.11.4.33543
Data Visibility :
Public

Validation Results

«
»

PixelMed Standard Report
XML

«
»
Validation Summary
1
4

DICOM SR ConformanceFAILED

Validation Summary
1
4
Constraint Description
Error: Template 10011 CTRadiationDose/[Row 1] CONTAINER (113701,DCM,"X-Ray Radiation Dose Report")/[Row 2] CODE (121058,DCM,"Procedure reported"): 1.1: /CONTAINER (113701,DCM,"X-Ray Radiation Dose Report")/CODE (121058,DCM,"Procedure reported"): Incorrect value - expected (P5-08000,SRT,"Computed Tomography X-Ray") - found (77477000,SCT,"Computed Tomography X-Ray")
Constraint Description
Found XRayRadiationDoseSR IOD
Constraint Description
Found Root Template TID_10011 (CTRadiationDose)
Constraint Description
Root Template Validation Complete
Constraint Description
IOD validation complete

File Content



Metadata
Metadata (XML)
Metadata (Tree)
Binary content

«
»

Unable to prettify DICOM: Null charset name
1